Malfunction
Summary report: N
PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)
MDR report key: 4151840
·
Received October 8, 2014
Report
- Report Number
- 1031452-2014-12659
- Event Type
- Malfunction
- Date Received
- October 8, 2014
- Report Date
- September 22, 2014
- Manufacturer
- INVACARE FLORIDA OPERATIONS
- Product Code
- CAW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Description of Event or Problem · 1
PER COMPLAINANT, ALARMING WITH LOW O2. INDICATION IS THAT SIEVE BED FILTERS (BOTTOM) ARE NOT SEATED PROPERLY ALLOWING SIEVE MATERIAL TO BECOME PULVERIZED AND BYPASS FILTER. SIEVE DUST IS MIGRATING INTO THE MANIFOLD VALVE AND MUFFLER. REFER TO (B)(4). SIEVE BEDS HAVE ALREADY BLOWN.