PENTA
Report
- Report Number
- 1627487-2014-12657
- Event Type
- Injury
- Date Received
- October 6, 2014
- Date of Event
- September 6, 2014
- Report Date
- September 8, 2014
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- GZB
- PMA / PMN Number
- P010032
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
METHOD: THE DEVICE HISTORY AND STERILIZATION RECORDS WERE REVIEWED. RESULTS: BATCH RECORDS REVIEW FOR 4606184 SHOWED NON-CONFORMANCES RELATED TO INSUFFICIENT MATERIAL FROM THE WAREHOUSE, LABELING DEFECTS AND FOREIGN MATERIAL. SOME UNITS WERE REPLACED AND THE DEVICE SHOWN TO BE MANUFACTURED PER SPECS. SOME UNITS WERE REPRINTED AND THE REMAINING UNITS WERE SCRAPPED. AS NONCONFORMING UNITS WERE NOT DISTRIBUTED, THESE NON-CONFORMANCE COULD NOT HAVE CONTRIBUTED TO THE PRODUCT FAILURE IDENTIFIED IN THE COMPLAINT FOR 'INFECTION'. CONCLUSION: THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
IT WAS REPORTED THE PT WENT TO THE EMERGENCY ROOM COMPLAINING OF PAIN AND SWELLING AT THE LEAD SITE AND WAS GIVEN PAIN MEDICATIONS SENT HOME. TWO DAYS LATER THE PT WAS SEEN BY THE IMPLANTING PHYSICIAN WHO NOTED SIGNS OF INFECTION. THE PT WAS ADMITTED TO THE HOSP AND IV ANTIBIOTICS WERE STARTED. THE PHYSICIAN FOUND THE LEAD SITE WAS INFECTED AND THE INFECTION HAD SPREAD TO THE IPG. BOTH DEVICES WERE EXPLANTED. CULTURES WERE TAKEN AND CONFIRMED A (B)(6) INFECTION. THE PT WAS SENT HOME AND WILL CONTINUE IV ANTIBIOTICS FOR SIX WEEKS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 623369 | PENTA | SCS LEAD | GZB | ST. JUDE MEDICAL - NEUROMODULATION | 3228 | 4606184 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR | Hospitalization | IMPLANT DATE:| SCS IPG, MODEL 3788, |