PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 4151801
·
Received October 8, 2014
Report
- Report Number
- 3004209178-2014-92470
- Event Type
- Malfunction
- Date Received
- October 8, 2014
- Date of Event
- July 13, 2014
- Report Date
- July 13, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
INSULIN PUMP RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM NOTED. INSULIN PUMP RECEIVED WITH MINOR SCRATCHES ON LCD WINDOW, CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS AND MISSING END CAP STICKER.
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Description of Event or Problem · 1
IT IS REPORTED THAT THE CUSTOMER HAS A BUTTON ERROR ON THE INSULIN PUMP. THEIR BLOOD GLUCOSE LEVEL WAS 124MG/DL. THE CUSTOMER STATES SHE TRIED TO CLEAR IT BUT CANNOT. TROUBLESHOOTING WAS DONE. NO FURTHER INFORMATION AVAILABLE.