EON
Report
- Report Number
- 1627487-2014-12659
- Event Type
- Injury
- Date Received
- October 6, 2014
- Date of Event
- September 9, 2014
- Report Date
- September 9, 2014
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- GZB
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS MODEL NUMBER WAS INCLUDED IN A FIELD CORRECTION. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 1 OF 2. REF MFR REPORT #1627487-2014-12660. IT WAS REPORTED THE PT EXPERIENCES HEATING AT THE IPG POCKET SITE WHEN CHARGING. THE PT ALSO REPORTED DARK MARKS OVER THE IPG. THE PHYSICIAN DOES NOT BELIEVE THE MARKS ARE RELATED TO THE SCS SYSTEM. A REPLACEMENT CHARGER WAS SENT TO THE PT. ON (B)(4)2012 ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 623368 | EON | SCS IPG | GZB | ST. JUDE MEDICAL - NEUROMODULATION | 3716 | 97739 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR | Other | IMPLANT DATE:| SCS LEAD, MODEL 3214, |