FDA Adverse Event
Malfunction
Summary report: N
ACCENT DR RF
MDR report key: 4151777
·
Received January 13, 2014
Report
- Report Number
- 2017865-2014-06364
- Event Type
- Malfunction
- Date Received
- January 13, 2014
- Date of Event
- March 30, 2012
- Manufacturer
- ST. JUDE MEDICAL, INC, CRMD
- Product Code
- DXY
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. THIS HISTORICAL COMPLAINT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF COMPLAINT FILES IN RESPONSE TO A RECENT FDA INSPECTION. THERE IS NO CHANGE TO THE ACTUAL PERFORMANCE OF THE PRODUCT AND THIS REPORT ONLY REPRESENTS AN ENHANCEMENT TO THE REPORTING CRITERIA GOING FORWARD. EVALUATION DESCRIPTION: THE REPORTED FIELD EVENT OF SET SCREW FAILURE COULD BE CONFIRMED IN THE LABORATORY. ANALYSIS FOUND MEDICAL ADHESIVE INSIDE BOTH SET SCREWS INSETS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE SET SCREW FAILED TO WORK PROPERLY. THE PACE MAKER WAS REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 27781 | ACCENT DR RF | IMPLANTABLE PACEMAKER PULSE GENERATOR, DXY | DXY | ST. JUDE MEDICAL, INC, CRMD | PM2210 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR |