FDA Adverse Event Injury Summary report: N

CD HORIZON SPINAL SYSTEM

MDR report key: 4151659 · Received October 8, 2014

Report

Report Number
1030489-2014-03911
Event Type
Injury
Date Received
October 8, 2014
Date of Event
August 1, 2014
Report Date
September 9, 2014
Manufacturer
MDT SOFAMOR DANEK PUERTO RICO MFG
Product Code
KWP
PMA / PMN Number
K052187
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PR, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

CORRECTION: SPECIFIC LOCATION WITHIN CONSTRUCT UNKNOWN. THREADS DO NOT SHOW EVIDENCE OF CROSS THREADING. MICROSCOPICALLY EXAMINED SET SCREW ROD INTERFACE NODE AND SURFACE; SET SCREW NODE HEIGHT AND MORPHOLOGY OF DEFORMATION ARE ATYPICAL OF FULL TIGHTENING. NODE DEFORMATION HEIGHT (@ CENTER) IDENTIFIED TO BE SIGNIFICANTLY HIGHER THAN A BENCH COMPARISON SAMPLE. MICROSCOPICALLY EXAMINED MAS HEAD; ASYMMETRICAL CROWN WITNESS MARKS NOTED ON THE SURFACE MAS, CONSISTENT WITH ANGULATED ROD AT FINAL TIGHTENING. THE ABOVE OBSERVATIONS ARE CONSISTENT WITH INCOMPLETE SEATING OF THE ROD AT FINAL TIGHTENING OF THE SET SCREW.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS NOT RETURNED FOR EVALUATION, HOWEVER FILMS WERE SUPPLIED FOR REVIEW WHICH FOUND: THREE VIEWS ARE PROVIDED AN AP AND LATERAL X-RAY AND A LATERAL REVERSE IMAGE LATERAL OF A LUMBAR SPINE AFTER SURGICAL TREATMENT FOR AN L2 FRACTURE. THESE HAVE APPARENTLY BEEN TAKEN AFTER THE FIRST REVISION AS THESE ARE LEGACY SCREWS AND RODS AND NOT LONGITUDE. THE FRACTURE APPEARS TO BE ABOUT 20% COMPRESSED ANTERIORLY WITH EXAGGERATED KYPHOSIS THROUGH THE LEVEL. PEDICLE SCREWS EXTEND FROM T12, L1, L3 AND L4. MINIMAL ROD CONTOURING IS PRESENT. THE SET SCREWS ON L3 AND L4 HAVE BACKED OUT ALLOWING SOME RECURRENCE OF PATHOLOGIC KYPHOSIS. NO SPECIFIC REASON FOR SET SCREW BACK OUT IS SEEN. PSEUDO ARTHRITIS CAN LEAD TO METAL FAILURE AND SCREW BACK OUT, HOWEVER IMPLANTS ARE DESIGNED TO SUPPORT FUSION FOR LONGER THAN 6 MONTHS IF THE PATIENT'S ACTIVITY LEVELS WERE APPROPRIATELY CONTROLLED. A 4TH IMAGE IS A CORONAL CT RECONSTRUCTION OF THE SAME T12 TO L4 CONSTRUCT. IT VERIFIES THAT THE PEDICLE SCREWS ARE WELL POSITIONED WITHIN THE PEDICLES AND THAT THERE IS A SAGITTAL SPLIT OF THE L2 FRACTURE DOCUMENTING IT AS A TWO COLUMN INJURY OR BURST FRACTURE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION SURGERY TO IMPLANT NEW HARDWARE AT T12-L4 TO TREAT A L2 FRACTURE. IT WAS REPORTED THAT 6 MONTHS POST-OP THE PATIENT EXPERIENCED PAIN IN THE LOWER BACK AND 7 MONTHS POST-OP THE PATIENT WENT IN TO SEE THE SURGEON. DURING THE VISIT THE PATIENT UNDERWENT A MRI AND CT WHICH SHOWED NON-UNION AND THAT THE LOWER 3 SET SCREWS HAD DISLODGES. THE SET SCREWS MIGRATION ALLOWED THE ROD TO MOVE AND PIERCE INTO THE PATIENT¿S SKIN. THE PATIENT UNDERWENT A REVISION TO REMOVE THE IMPLANTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
633189 CD HORIZON SPINAL SYSTEM APPLIANCE, FIXATION, SPINAL INTERLAMINAL KWP MDT SOFAMOR DANEK PUERTO RICO MFG NA H13J4858

Patients

Seq Age Sex Outcome Treatment
1 00032 YR Hospitalization| R SCREWS, RODS, BMP