PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4151594 · Received October 8, 2014

Report

Report Number
2032227-2014-34670
Event Type
Malfunction
Date Received
October 8, 2014
Date of Event
September 5, 2014
Report Date
September 8, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

INSULIN PUMP RECEIVED WITH INTERMITTENT BUTTON RESPONSE, DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM. INSULIN PUMP RECEIVED WITH CRACKED CASE ON DISPLAY WINDOW CORNER AND CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

IT IS REPORTED THAT THE PATIENT'S SENSOR INSULIN PUMP DISPLAYED A BUTTON ERROR. THE PATIENT'S BLOOD GLUCOSE LEVEL WAS 100MG/DL. TROUBLE SHOOTING WAS CONDUCTED AND WE EXPLAINED THE POSSIBLE CAUSES FOR THE ALARM. DEVICE IS BEING RETURNED. NO FURTHER INFORMATION AVAILABLE.