PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-754LCAP
Report
- Report Number
- 2032227-2014-34666
- Event Type
- Malfunction
- Date Received
- October 8, 2014
- Date of Event
- September 7, 2014
- Report Date
- September 8, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IS
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
INSULIN PUMP RECEIVED WITH BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM. INSULIN PUMP RECEIVED WITH MINOR SCRATCHES ON DISPLAY WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP.
IT IS REPORTED THAT THE PATIENT'S SENSOR INSULIN PUMP DISPLAYED A BUTTON ERROR. THE PATIENT'S BLOOD GLUCOSE LEVEL WAS 200MG/DL. TROUBLE SHOOTING WAS CONDUCTED AND WE EXPLAINED THE POSSIBLE CAUSES FOR THE ALARM. DEVICE IS BEING RETURNED. NO FURTHER INFORMATION AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 633516 | PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-754LCAP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-754LCAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 114 YR |