PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-34649
- Event Type
- Malfunction
- Date Received
- October 8, 2014
- Date of Event
- September 4, 2014
- Report Date
- September 8, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. INSULIN PUMP RECEIVED WITH CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON LCD WINDOW AND SCRATCHED RESERVOIR TUBE WINDOW.
IT WAS REPORTED BY THE CUSTOMER'S FATHER THAT THE CUSTOMER HAD A KEYPAD ANOMALY ON THE INSULIN PUMP. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 357 MG/DL. CUSTOMER STATED THAT THE UP KEY ARROW WAS NOT WORKING. CUSTOMER WAS ABLE TO GIVE INSULIN WITH THE PUMP. INSULIN PUMP WILL NEED TO BE REPLACED. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE PUMP AND REVERT TO A BACK-UP PLAN PER HEALTH CARE PROFESSIONAL'S INSTRUCTION. NO ADDITIONAL INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 633513 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-523LNAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 14 YR |