FDA Adverse Event Malfunction Summary report: N

10.5CM ANGLE ATTACHMENT

MDR report key: 4151514 · Received October 8, 2014

Report

Report Number
1045834-2014-13659
Event Type
Malfunction
Date Received
October 8, 2014
Report Date
September 24, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
PK011444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. (B)(4) EVALUATED THE DEVICE AND OBSERVED THAT THE DEVICE FAILED THE CUTTER INSERTION ASSESSMENT. IT WAS DETERMINED THAT THE BEARINGS WERE WORN AND WERE NO LONGER IN AXIAL ALIGNMENT. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE OF THE BEARINGS BEING OUT OF ALIGNMENT WAS DETERMINED TO BE NORMAL WEAR FROM USE AND SERVICING. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE WHEN THE CUTTER DEVICE WAS INSERTED, THE ATTACHMENT DEVICE DID NOT SPIN. IT WAS NOT REPORTED THAT THE DEVICE WAS USED IN SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
633448 10.5CM ANGLE ATTACHMENT MOTOR, DRILL, ELECTRIC - ATTACHMENT HBC DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1