FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP

MDR report key: 4151499 · Received October 8, 2014

Report

Report Number
2032227-2014-34624
Event Type
Malfunction
Date Received
October 8, 2014
Date of Event
September 8, 2014
Report Date
September 8, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
PMA / PMN Number
P120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4).

Additional Manufacturer Narrative · 1

A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS AND A CRACKED CASE NEAR THE DISPLAY WINDOW CORNERS WERE NOTED DURING THE VISUAL INSPECTION. ALL BUTTONS FUNCTIONED PROPERLY. HOWEVER, CORRODED KEYPAD TRACES WERE NOTED. THE INSULIN PUMP PASSED THE PRIME, DISPLACEMENT, BASIC OCCLUSION, OCCLUSION AND EXCESSIVE NO DELIVERY ALARM TESTS.

Description of Event or Problem · 1

CUSTOMER REPORTED A BUTTON ERROR ALARM AFTER HER BUTTONS BECAME FROZEN. SHE ALSO STATED THE ISSUE OCCURRED WHILE SHE WAS ATTEMPTING A BOLUS. THE CUSTOMER'S BLOOD GLUCOSE WAS 240 MG/DL. THE CUSTOMER COULD NOT RECALL ANY SIGNIFICANT EVENTS LEADING TO THE ALARM. CUSTOMER WAS ADVISED TO DISCONNECT FROM THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. NO ADDITIONAL INFORMATION IS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
633443 530G INSULIN PUMP OZO OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-751NAH

Patients

Seq Age Sex Outcome Treatment
1 32 YR