FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4151491 · Received October 8, 2014

Report

Report Number
2032227-2014-34617
Event Type
Malfunction
Date Received
October 8, 2014
Date of Event
September 8, 2014
Report Date
September 8, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM WAS NOTED DURING TESTING. NO MOISTURE DAMAGE WAS FOUND ON THE ELECTRONICS, MOTOR, BATTERY TUBE OR VIBRATOR ASSEMBLY DURING THE VISUAL INSPECTION. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, A CRACKED CASE AT A DISPLAY WINDOW CORNER, MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED DISPLAY WINDOW, A SCRATCHED RESERVOIR TUBE WINDOW AND A CRACKED BELT CLIP SLOT.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THIS MDR RELATED TO THE (B)(4) MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED A BUTTON ERROR ALARM. SHE STATED THE ALARM OCCURRED WHILE SHE WAS ATTEMPTING TO GIVE HERSELF INSULIN. THE CUSTOMER'S BLOOD GLUCOSE WAS UNKNOWN. THE CUSTOMER STATED HER INSULIN PUMP MAY HAVE BEEN EXPOSED TO SWEAT. SHE STATED THERE WAS FLUID UNDER HER LCD DISPLAY. THE CUSTOMER STATED SHE DID NOT DROP HER INSULIN PUMP. SHE ALSO REPORTED SEVERAL CRACKS AROUND THE PLASTIC CASING OF HER INSULIN PUMP. CUSTOMER WAS ADVISED TO DISCONNECT FROM THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. NO ADDITIONAL INFORMATION IS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
633241 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAL

Patients

Seq Age Sex Outcome Treatment
1 24 YR