FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4151488 · Received October 8, 2014

Report

Report Number
2032227-2014-34537
Event Type
Malfunction
Date Received
October 8, 2014
Date of Event
September 8, 2014
Report Date
September 8, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM NOTED. THE INSULIN PUMP HAS MINOR SCRATCHED DISPLAY WINDOW, CRACKED RESERVOIR TUBE LIP AND MISSING THE END CAP STICKER.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. CUSTOMER'S BLOOD GLUCOSE READING WAS 216 MG/DL. NO SIGNIFICANT EVENTS LEADING TO THE ALARM WERE OBSERVED. ADVISED DISCONTINUATION OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
633240 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-554WWS

Patients

Seq Age Sex Outcome Treatment
1