FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4151451 · Received October 8, 2014

Report

Report Number
2032227-2014-34600
Event Type
Malfunction
Date Received
October 8, 2014
Date of Event
September 5, 2014
Report Date
September 8, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP RECEIVED WITH AN INTERMITTENT BUTTON RESPONSE DUET O CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM. INSULIN PUMP RECEIVED WITH MINOR SCRATCHES DISPLAY WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, MISSING END CAP STICKER AND CRACKED LCD WINDOW.

Additional Manufacturer Narrative · 1

(B)(4).CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT IS REPORTED THAT THE PATIENT'S SENSOR INSULIN PUMP DISPLAYED A BUTTON ERROR. THE PATIENT'S BLOOD GLUCOSE LEVEL WAS 167MG/DL. TROUBLE SHOOTING WAS CONDUCTED AND WE EXPLAINED THE POSSIBLE CAUSES FOR THE ALARM. DEVICE IS BEING RETURNED. NO FURTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
632808 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-754CAB

Patients

Seq Age Sex Outcome Treatment
1 114 YR