FDA Adverse Event
Malfunction
Summary report: N
PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)
MDR report key: 4151247
·
Received October 8, 2014
Report
- Report Number
- 1031452-2014-12632
- Event Type
- Malfunction
- Date Received
- October 8, 2014
- Report Date
- September 16, 2014
- Manufacturer
- INVACARE FLORIDA OPERATIONS
- Product Code
- CAW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Description of Event or Problem · 1
PER INDEPENDENT REPAIR CENTER STATEMENT THE UNIT WAS ALARMING OR RED LIGHT. THE KEY FAILURE WAS THE 4 WAY VALVE NOT SWITCHING. ADDITIONAL MALFUNCTIONS INCLUDE THE VALVE OPERATOR WAS STUCK, THE METAL CLAMPS WERE LEAKING, AND THE SIEVE BEDS HAD LOW O2.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 632541 | PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) | 868.5440 | CAW | INVACARE FLORIDA OPERATIONS | IRC5PO2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |