ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-28594
- Event Type
- Malfunction
- Date Received
- October 8, 2014
- Report Date
- September 27, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- PATIENT
Narratives
FOLLOW-UP #1: DATE OF SUBMISSION 12/08/2014. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 11/17/2014 WITH THE FOLLOWING FINDINGS: THERE WERE NO ¿EMPTY CARTRIDGE¿ ALARMS OBSERVED IN THE BLACK BOX HISTORY. ALL CARTRIDGE CHANGES IN THE BLACK BOX OCCURRED BEFORE THE CARTRIDGE WAS EMPTY. AN ¿EMPTY CARTRIDGE¿ ALARM WAS DUPLICATED DURING TESTING. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS AND ALLEGED THAT THE PUMP EMITTED AN EMPTY CARTRIDGE ALARM THAT WAS NOT RECORDED IN THE ALARM HISTORY. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 633154 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 11 YR |