ALTRUA
Report
- Report Number
- 2124215-2014-16860
- Event Type
- Malfunction
- Date Received
- October 8, 2014
- Date of Event
- January 1, 2011
- Report Date
- August 30, 2018
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWP
- PMA / PMN Number
- P970003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. PIN GAUGE TESTING, DESIGNED TO VERIFY PROPER PORT DIMENSIONS, WAS COMPLETED. EACH PORT MEASURED AS EXPECTED. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING, AND SENSING FUNCTIONS WERE TESTED. IMPEDANCE TESTING WAS COMPLETED AND ALL MEASUREMENTS WERE WITHIN NORMAL LIMITS. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED. ANALYSIS DID NOT IDENTIFY ANY DEVICE CHARACTERISTICS THAT WOULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED CLINICAL OBSERVATIONS.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE PACEMAKER WITH THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED GRADUAL INCREASE IN PACING LEAD IMPEDANCE MEASUREMENT, THOUGH STILL WITHIN NORMAL RANGE. A FLUCTUATION IN PACING THRESHOLD WAS ALSO NOTED. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) SUGGESTED X-RAY, ISOMETRICS AND EVALUATION OF THE SYSTEM IN UNIPOLAR CONFIGURATION; UNABLE TO RULE OUT A PROXIMAL CONDUCTOR ISSUE. TS THEN PROVIDED LEAD IMPEDANCE RANGE. ADDITIONAL INFORMATION WAS RECEIVED INDICATING THAT THREE MONTHS LATER THIS SYSTEM EXHIBITED HIGH, OUT-OF-RANGE (OOR) PACING IMPEDANCE MEASUREMENT OF MORE THAN 2,000 OHMS. TS DISCUSSED TROUBLESHOOTING OPTIONS. SUBSEQUENTLY, THE SYSTEM CONFIGURATION WAS CHANGED TO UNIPOLAR AND IMPEDANCE MEASUREMENT DECREASED TO 860 OHMS. AT THIS TIME, NO FURTHER INTERVENTION WAS PERFORMED. THE SYSTEM REMAINS IN SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 632523 | ALTRUA | IMPLANTABLE PULSE GENERATOR | LWP | GUIDANT CRM CLONMEL IRELAND | S603 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 YR | 4088| 4087| S603 |