FDA Adverse Event Malfunction Summary report: N

INSIGNIA

MDR report key: 4150842 · Received October 8, 2014

Report

Report Number
2124215-2014-17980
Event Type
Malfunction
Date Received
October 8, 2014
Date of Event
July 7, 2014
Report Date
September 3, 2014
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, DETAILED MECHANICAL AND ELECTRICAL TESTING WAS PERFORMED ON THE DEVICE. THE DEVICE BATTERY STATUS WAS EOL. THE DEVICE FUNCTIONED NORMALLY THROUGHOUT TESTING. LABORATORY ANALYSIS DETERMINED THAT THIS DEVICE EXPERIENCED NORMAL BATTERY DEPLETION; HOWEVER, THE ESTIMATED LONGEVITY REMAINING VALUE APPEARED TO DECREASE MORE QUICKLY THAN EXPECTED BETWEEN ROUTINE FOLLOW-UPS. FACTORS INFLUENCING THE ESTIMATED LONGEVITY REMAINING CALCULATION INCLUDE PACING RATE, AMPLITUDE, PULSE-WIDTH AND LEAD IMPEDANCE. ANY (EVEN SLIGHT) CHANGES IN THESE FACTORS WILL IMPACT THE BATTERY CONSUMPTION CALCULATION AND THEREFORE THE REMAINING LONGEVITY ESTIMATE. PLEASE NOTE THAT, DESPITE THE DROP IN ESTIMATED LONGEVITY REMAINING, THE ACTUAL BATTERY CONDITION DID NOT CHANGE SIGNIFICANTLY BETWEEN FOLLOW-UPS. IN SUMMARY, IT WAS DETERMINED THAT THIS DEVICE EXPERIENCED NORMAL BATTERY DEPLETION, BUT DECLARED EOL EARLIER THAN PREVIOUSLY ESTIMATED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PACEMAKER SHOWED MAGNET RATE OF NINETY AND HALF A YEAR REMAINING BATTERY LIFE. HOWEVER, THE PACEMAKER DECLARED END OF LIFE (EOL) FOUR MONTHS AFTER. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) DISCUSSED THERAPY MAY NOT BE GUARANTEED AND ADVISED TO HAVE A DEVICE CHANGEOUT. THIS PACEMAKER WAS EXPLANTED AND A NEW PACEMAKER WAS IMPLANTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
632746 INSIGNIA IMPLANTABLE PULSE GENERATOR LWP GUIDANT CRM CLONMEL IRELAND 1298

Patients

Seq Age Sex Outcome Treatment
1 78 YR 4088| 4087| 1298