FDA Adverse Event
Malfunction
Summary report: N
THE EMAX 2 PLUS SYSTEM
MDR report key: 4150826
·
Received October 8, 2014
Report
- Report Number
- 1045834-2014-13362
- Event Type
- Malfunction
- Date Received
- October 8, 2014
- Report Date
- February 14, 2013
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HBC
- PMA / PMN Number
- PK080802
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES OUS SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. DURING PRE-REPAIR ASSESSMENT PERFORMED BY TECHNICIAN, IT WAS NOTED THAT THE TENSIONING PIN FROM THE ROTATING JOINT IS MISSING AND THE WAREHOUSE FROM THE TOOL LOCK IS NOISY DUE TO NORMAL WEAR. DEVICE WAS INSPECTED AND REPAIRED.
Description of Event or Problem · 1
CUSTOMER REPORTED THAT THE DEVICE WAS MAKING NOISE AND CREATING HEAT. THIS IS 1 OF 1 REPORT FOR EVENT # (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 632427 | THE EMAX 2 PLUS SYSTEM | MOTOR, DRILL, ELECTRIC | HBC | DEPUY SYNTHES POWER TOOLS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |