FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4150302 · Received October 7, 2014

Report

Report Number
2032227-2014-34300
Event Type
Malfunction
Date Received
October 7, 2014
Date of Event
March 30, 2014
Report Date
March 30, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4).A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS. THE INSULIN PUMP FUNCTION PROPERLY DURING REWIND AND BASIC OCCLUSION, OCCLUSION, PRIME, THE EXCESSIVE NO DELIVERY AND DISPLACEMENT TEST. NO EXCESSIVE NO DELIVERY ALARM NOTED. THE DEVICE WAS RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP AND CRACKED BATTERY TUBE THREADS.

Description of Event or Problem · 1

CUSTOMER REPORTED A NO DELIVERY ALARM DURING A BASAL DELIVERY. CUSTOMER'S BLOOD GLUCOSE WAS 124 MG/DL. SHE STATED THE ISSUE WAS NOT RESOLVED WITH AN INFUSION SET, SITE, OR RESERVOIR CHANGE. THE CUSTOMER WAS ADVISED TO DISCONNECT FROM THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
629751 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 35 YR