FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4150297 · Received October 7, 2014

Report

Report Number
2032227-2014-34299
Event Type
Malfunction
Date Received
October 7, 2014
Date of Event
November 4, 2013
Report Date
November 4, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).THE INSULIN PUMP WAS RECEIVED WITH A MOTOR ERROR ALARM DURING REWIND DUE TO CORRODED MOTOR HOMES WITCH. THE DEVICE WAS RECEIVED WITH CRACKED RESERVOIR TUBE LIP. THE DEVICE WAS RECEIVED WITH CRACKED BELT CLIP SLOT.THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

IT WAS REPORTED THE CUSTOMER RECEIVED A MOTOR ERROR ALARM. NO ADDITIONAL INFORMATION PROVIDED.