FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4150291 · Received October 7, 2014

Report

Report Number
2032227-2014-34963
Event Type
Malfunction
Date Received
October 7, 2014
Date of Event
September 8, 2014
Report Date
September 8, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP FUNCTIONED PROPERLY DURING THE REWIND, BASIC OCCLUSION, PRIME, EXCESSIVE NO DELIVERY ALARM AND DISPLACEMENT TESTS. NO REWIND ANOMALY OR PRIME ANOMALY WAS NOTED. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.(B)(4)

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP WAS STUCK IN REWIND LOOP. THE CUSTOMER'S BLOOD GLUCOSE WAS 120 MG/DL. THE CUSTOMER WAS ADVISED TO REVERT TO A BACK-UP PLAN. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
629169 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723LNAS

Patients

Seq Age Sex Outcome Treatment
1 55 YR