G4 PLATINUM PEDIATRIC CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2014-25597
- Event Type
- Malfunction
- Date Received
- October 7, 2014
- Date of Event
- September 10, 2014
- Report Date
- September 10, 2014
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
(B)(4). THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE REPORTED FAULT COULD NOT BE REPRODUCED AND THERE WAS NO FAILURE DETECTED. THE DEVICE WAS DETERMINED TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION. AN ALARM TEST AND AN INTERIOR INSPECTION WAS PERFORMED AND THE DEVICE FAILED, CONFIRMING THE REPORTED EVENT OF LOW AUDIO OUTPUT. THE ROOT CAUSE WAS DETERMINED TO BE AN ASSEMBLY ERROR.
(B)(4).
PATIENT'S FATHER CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6)2014 TO CLAIM LOW AUDIO OUTPUT ON (B)(6)2014. AT THE TIME CONTACT WITH DEXCOM TECHNICAL SUPPORT, NO MEDICAL INTERVENTION OR INJURIES WERE REPORTED. ADDITIONALLY, AT THE TIME OF CONTACT, DEXCOM TECHNICAL SUPPORT ADVISED PATIENT'S FATHER TO TEST THE ALERT FUNCTIONALITY AND PATIENT'S FATHER REPORTED ALERTS WERE FUNCTIONAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 627846 | G4 PLATINUM PEDIATRIC CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | MT22430-BLU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 7 YR |