FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4150079 · Received October 7, 2014

Report

Report Number
2032227-2014-34831
Event Type
Malfunction
Date Received
October 7, 2014
Date of Event
September 8, 2014
Report Date
September 8, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.(B)(4)

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE DISPLACEMENT TEST AND THE SELF TEST. NO UNEXPECTED SOFTWARE ERROR ALARMS WERE NOTED. HOWEVER, CORRODED ELECTRONIC ASSEMBLY AND MOTOR ASSEMBLY WAS NOTED DURING THE VISUAL INSPECTION. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW, CRACKED RESERVOIR TUBE LIP AND CRACKED BATTERY TUBE THREADS.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING AN ERROR ALARM ON THE INSULIN PUMP. CUSTOMER REPORTED THAT CONDENSATION WAS VISIBLE UNDER THE DISPLAY WINDOW. CUSTOMER STATED THAT SHE HAD RECENTLY JUMPED INTO A SWIMMING POOL WHILE WEARING THE INSULIN PUMP. BLOOD GLUCOSE LEVEL WAS NOT PROVIDED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
627143 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAL

Patients

Seq Age Sex Outcome Treatment
1 66 YR