PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 4150044
·
Received October 7, 2014
Report
- Report Number
- 2032227-2014-34810
- Event Type
- Malfunction
- Date Received
- October 7, 2014
- Date of Event
- September 8, 2014
- Report Date
- September 8, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).
Additional Manufacturer Narrative · 1
INSULIN PUMP ALARMED DURING SELF TEST DUE TO FAULTY CONNECTOR ON REMOTE FREQUENCY BOARD. INSULIN PUMP RECEIVED WITH MINOR SCRATCHES ON LCD WINDOW, CRACKED RESERVOIR TUBE LIP, BROKEN BELT CLIP SLOT AND CRACKED BATTERY TUBE THREADS.
Description of Event or Problem · 1
IT WAS REPORTED BY THE CUSTOMER'S MOTHER, THAT THE CUSTOMER RECEIVED A FAILED SELFTEST ALARM. CUSTOMER'S BLOOD GLUCOSE LEVEL AT THE TIME 340MG/DL. ADVISED INSULIN PUMP WOULD BE REPLACED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 626867 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-523LNAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 7 YR |