FDA Adverse Event Malfunction Summary report: N

SUR-FIT NATURA UROSTOMY POUCH WITH ACCUSEAL TAP W/VALVE TRANSPARENT STANDARD W/

MDR report key: 4148476 · Received October 2, 2014

Report

Report Number
9618003-2014-11553
Event Type
Malfunction
Date Received
October 2, 2014
Date of Event
April 16, 2013
Report Date
April 16, 2013
Manufacturer
CONVATEC DOM. REPUBLIS, INC.
Product Code
EXH
PMA / PMN Number
K840166
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

AN INVESTIGATION INTO AN INCREASE OF COMPLAINTS REGARDING A BLOCKED TAP WAS PERFORMED. ANALYSIS OF THOSE COMPLAINTS FOUND THE ROOT CAUSE OF THE EVENT WAS AN EXCESS AMOUNT OF SILICONE APPLIED TO THE TAP DURING THE MANUFACTURING PROCESS. CONVATEC IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO THE FDA 483 ISSUED APRIL 16, 2014. DEVICE(S) LISTED IN THIS REPORT IS (ARE) USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR.

Description of Event or Problem · 1

IT WAS REPORTED THE END USER COULD NOT DRAIN URINE FORM THE POUCH VIA THE TAP. NO PT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.