Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/24/2014 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THE LABEL WAS TORN. THE DISPLAY HAD A PINK CONTRAST. A DISPLAY LENS LEAK WAS FOUND DURING LEAK TEST. MOISTURE WAS SEEN ON THE DISPLAY AND INSIDE THE PUMP BUT NOT IN THE BATTERY COMPARTMENT. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. EVALUATION REVEALED THAT THE DISPLAY HAD A PINK CONTRAST. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 09/24/2014.