FDA Adverse Event Malfunction Summary report: N

2.0MM DRILL BIT W/DEPTH MARK QC/110MM

MDR report key: 4145805 · Received October 6, 2014

Report

Report Number
3000270450-2014-10116
Event Type
Malfunction
Date Received
October 6, 2014
Date of Event
September 10, 2014
Report Date
September 11, 2014
Manufacturer
SYNTHES SELZACH
Product Code
GFG
PMA / PMN Number
PEXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HU
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. NO NONCONFORMANCE REPORTS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. THE VISUAL INSPECTION OF THE RETURNED DEVICE BY THE COMPLAINT HANDLING UNIT REVEALED THE INSTRUMENTS ARE BROKEN AND THE REMAINING RETURNED SHAFTS WITH PARTIAL EDGES ARE BENT. EACH OF THE TWO DRILL BITS WERE RETURNED INCOMPLETE. THE SECOND PART FROM BOTH WITH EDGES AND DISTAL TIPS ARE MISSING AND THEREFORE COULD NOT BE EXAMINED. INVESTIGATION OF DOCUMENTATION FOR MANUFACTURING AND MATERIAL SHOWS THAT THE INSTRUMENTS PRODUCED IN 2012 PRINT SPECIFICATION. NO FAILURE IN MATERIAL OR MANUFACTURING COULD BE DETECTED THE COMPLAINT CONDITION IS LIKELY THE RESULT THAT THE INSTRUMENTS GOT STUCK EITHER IN THE BONE OR IN THE DRILL SLEEVE AND AS THE USER TURNED WITH HIGH FORCE, THE INSTRUMENTS GOT DAMAGED. WE RECOMMEND NOT TURNING THE DRILL BITS DURING INSERTION INTO THE DRILL SLEEVE IN ORDER TO PREVENT JAMMING IN IT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THE DRILL BITS WERE BROKEN. THERE WASN¿T ANY IMPACT ON THE PROCEDURE, THE OPERATION WAS FINISHED SUCCESSFULLY. THIS IS REPORT 2 OF 2 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
622316 2.0MM DRILL BIT W/DEPTH MARK QC/110MM BIT, SURGICAL GFG SYNTHES SELZACH 421160

Patients

Seq Age Sex Outcome Treatment
1