FDA Adverse Event Malfunction Summary report: N

GB SCD EXPRESS COMP SYSTEM X1

MDR report key: 4143599 · Received October 6, 2014

Report

Report Number
3006451981-2014-00745
Event Type
Malfunction
Date Received
October 6, 2014
Date of Event
September 18, 2014
Report Date
September 18, 2014
Manufacturer
COVIDIEN
Product Code
JOW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ONE SCD EXPRESS COMPRESSION SYSTEM WAS RETURNED TO AN INTERNATIONAL SERVICE CENTER FOR THE REPORTED CONDITION OF; POWER CORD EXPOSED DUE TO BEING CUT BY THE BEDSIDE HANDLES. THE INTERNATIONAL SERVICE NOTES INDICATE THAT THE POWER CORD IS CUT, CONFIRMING THE REPORTED CONDITION. THE INNER INSULATION WAS NOT EXPOSED SO THIS DID NOT POSE AN ELECTRICAL SAFETY HAZARD TO THE USER OR CLINICAL STAFF. THE ROOT CAUSE CAN BE ATTRIBUTED TO ROUGH HANDLING. POWER CORDS PERIODICALLY REQUIRE REPLACEMENT DUE TO AGE, USAGE AND USER DAMAGE. THE SERVICE MANUAL INSTRUCTS THE USER TO PERIODICALLY INSPECT THE POWER CORDS RESISTANCE TO ENSURE ITS ELECTRICAL SAFETY. THE POWER CORD NEEDS TO BE REPLACED TO CORRECT THE PROBLEM. PRODUCT SCD EXPRESS WAS MANUFACTURED IN 2011. ALL DEVICE HISTORY RECORDS ARE REVIEWED FOR QUALITY INSPECTIONS AND PARAMETER COMPLIANCE PRIOR TO RELEASING THE PRODUCT FOR SHIPMENT. A REVIEW OF THE SERVICE HISTORY RECORDS INDICATES THERE ARE NO SERVICE HISTORIES RECORDED FOR THIS SYSTEM. THIS INFORMATION WILL BE UTILIZED FOR TRENDING PURPOSES TO DETERMINE THE NEED FOR CORRECTIVE ACTION.

Additional Manufacturer Narrative · 1

SUBMIT DATE: 10/06/2014.AN INVESTIGATION IS CURRENTLY UNDERWAY. UPON COMPLETION, THE RESULTS WILL BE FORWARDED.

Description of Event or Problem · 1

IT WAS REPORTED TO COVIDIEN ON (B)(6) 2014 THAT THE CUSTOMER HAD A POWER CORD EXPOSED DUE TO BEING CUT BY THE BEDSIDE HANDLES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
624882 GB SCD EXPRESS COMP SYSTEM X1 SCD CONTROLLER JOW COVIDIEN 95251S

Patients

Seq Age Sex Outcome Treatment
1