PROGRAMMING SOFTWARE
Report
- Report Number
- 1644487-2014-02556
- Event Type
- Malfunction
- Date Received
- October 6, 2014
- Date of Event
- September 8, 2014
- Report Date
- September 8, 2014
- Manufacturer
- CYBERONICS INC
- Product Code
- LYJ
- PMA / PMN Number
- P970003
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE MANUFACTURING RECORDS WERE REVIEWED. REVIEW OF MANUFACTURING RECORDS OF THE HANDHELD CONFIRMED ALL QUALITY TESTS WERE PASSED PRIOR TO DISTRIBUTION. DEVICE FAILURE OCCURRED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.
IT WAS REPORTED THAT THE PHYSICIAN'S HANDHELD WOULD FREEZE AT THE BLUE WINDOWS SCREEN AFTER RESET. TROUBLESHOOTING WAS PERFORMED, WHICH RESOLVED THE ISSUE. IT WAS LATER REPORTED THAT THE HANDHELD STOPPED WORKING AGAIN AND TROUBLESHOOTING WAS UNABLE TO RESOLVE THE ISSUE. IT WAS REPORTED THAT THE SCREEN LOCK WAS NOT ENGAGED AND THE SCREEN WAS CLEAN. IT WAS REPORTED THAT THE ALIGN SCREEN WOULD NOT WORK AND WHEN TAPPING ON THE SCREEN WITH THE STYLUS THERE WAS NO RESPONSE. A NEW PROGRAMMING TABLET WAS SENT TO THE PHYSICIAN AND THE HANDHELD AND WAND WERE RECEIVED FOR ANALYSIS. ANALYSIS OF THE WAND WAS COMPLETED ON (B)(4) 2014. THE PROGRAMMING WAND PERFORMED ACCORDING TO FUNCTIONAL SPECIFICATIONS. ANALYSIS OF THE HANDHELD WAS COMPLETED ON (B)(4) 2014. DURING THE ANALYSIS, IT WAS IDENTIFIED THAT THE HANDHELD WAS UNABLE TO ADVANCE PAST THE WELCOME SCREEN. THE CAUSE FOR THE DISPLAY ANOMALY IS ASSOCIATED WITH RESISTANCE VALUES BEING HIGHER THAN EXPECTED IN THE TOUCH SCREEN CIRCUITRY. ONCE THE DISPLAY WAS REPLACED WITH A KNOWN GOOD DISPLAY, NO FURTHER ANOMALIES ASSOCIATED WITH THE HANDHELD PERFORMANCE WERE IDENTIFIED DURING THE ANALYSIS. ANALYSIS OF THE FLASHCARD WAS COMPLETED ON (B)(4) 2014. NO ANOMALIES ASSOCIATED WITH FLASHCARD SOFTWARE OR DATABASES WERE IDENTIFIED DURING THE FLASHCARD ANALYSIS. THE FLASHCARD AND SOFTWARE PERFORMED ACCORDING TO FUNCTIONAL SPECIFICATIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 622332 | PROGRAMMING SOFTWARE | PROGRAMMING COMPUTER | LYJ | CYBERONICS INC | MODEL 250 | 1075854 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |