FDA Adverse Event Malfunction Summary report: N

CAD II

MDR report key: 4143373 · Received October 6, 2014

Report

Report Number
8030965-2014-01516
Event Type
Malfunction
Date Received
October 6, 2014
Report Date
November 1, 2012
Manufacturer
SYNTHES GMBH
Product Code
HWE
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO (B)(4) AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN, UNAVAILABLE OR UNCHANGED. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. MANUFACTURING DATE FOR THIS DEVICE IS UNAVAILABLE. DURING PRE-REPAIR ASSESSMENT PERFORMED BY TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED THE TRIGGER/SLIDER WAS BLOCKED. THIS WAS CONTRIBUTED TO NORMAL WEAR. UNIT WAS REPAIRED AND RETURNED TO THE CUSTOMER ON 11/01/2012. PLACEHOLDER

Description of Event or Problem · 1

CUSTOMER REPORTED THAT DEVICE KEPT TURNING EVEN WHEN FINGER WAS NOT ON THE TRIGGER. THIS IS REPORT 1 OF 1 FOR COMPLAINT #(B)(4).