ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-28365
- Event Type
- Malfunction
- Date Received
- October 6, 2014
- Report Date
- September 25, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PATIENT
Narratives
DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 11/22/2014 WITH THE FOLLOWING FINDINGS: THE PUMP WAS RETURNED WITH NO VISIBLE DAMAGE TO THE KEYPAD COVER. DURING TESTING, ALL OF THE KEYPAD BUTTONS WERE INTERMITTENTLY RESPONSIVE. THE KEYPAD COVER WAS REMOVED AND CONTAMINATION WAS FOUND UNDER THE KEYPAD BUTTON CONTACTS. UNRELATED TO THE INITIAL COMPLAINT, THE BATTERY COMPARTMENT WAS CRACKED.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THE REPORTER ALLEGED THAT THE DOWN ARROW KEYPAD BUTTON WAS UNDER RESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 622575 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 37 YR |