FDA Adverse Event Malfunction Summary report: N

OUTLOOK 200

MDR report key: 4143289 · Received September 26, 2014

Report

Report Number
4143289
Event Type
Malfunction
Date Received
September 26, 2014
Date of Event
September 15, 2014
Report Date
September 26, 2014
Manufacturer
B. BRAUN MEDICAL, INC.
Product Code
FRN
Product Problem
Yes
Report Source
User Facility report
Reporter Location
CA, US

Narratives

Description of Event or Problem · 1

PUMP READS "BAG EMPTY" AND ALARMS. REPORTED TO BIOMED. FOUND INLET RESTRICTION VALVE OUT OF ADJUSTMENT. CALIBRATED VALVE, FUNCTIONAL TEST COMPLETED AND PASSED. PUMP PUT BACK IN SERVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
600139 OUTLOOK 200 PUMP, INFUSION FRN B. BRAUN MEDICAL, INC. OUTLOOK 200 *

Patients

Seq Age Sex Outcome Treatment
1 *