FDA Adverse Event Malfunction Summary report: N

OUTLOOK 200

MDR report key: 4143285 · Received September 26, 2014

Report

Report Number
4143285
Event Type
Malfunction
Date Received
September 26, 2014
Date of Event
March 11, 2014
Report Date
September 26, 2014
Manufacturer
B BRAUN MEDICAL, INC.
Product Code
FRN
Product Problem
Yes
Report Source
User Facility report
Reporter Location
CA, US

Narratives

Description of Event or Problem · 1

REPORTED TO BIOMED THAT PUMP NOT WORKING. ENGINEER EVALUATED PUMP AND FOUND CONSTANT OCCLUSION ALARM; CLEANED AND LUBRICATED OUTLET RESTRICTION VALVE. FUNCTIONAL TEST COMPLETED AND PASSED. PUMP PUT BACK IN SERVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
599929 OUTLOOK 200 INFUSION PUMP FRN B BRAUN MEDICAL, INC. OUTLOOK 200 *

Patients

Seq Age Sex Outcome Treatment
1 *