FDA Adverse Event
Malfunction
Summary report: N
OUTLOOK 200
MDR report key: 4143285
·
Received September 26, 2014
Report
- Report Number
- 4143285
- Event Type
- Malfunction
- Date Received
- September 26, 2014
- Date of Event
- March 11, 2014
- Report Date
- September 26, 2014
- Manufacturer
- B BRAUN MEDICAL, INC.
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- CA, US
Narratives
Description of Event or Problem · 1
REPORTED TO BIOMED THAT PUMP NOT WORKING. ENGINEER EVALUATED PUMP AND FOUND CONSTANT OCCLUSION ALARM; CLEANED AND LUBRICATED OUTLET RESTRICTION VALVE. FUNCTIONAL TEST COMPLETED AND PASSED. PUMP PUT BACK IN SERVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 599929 | OUTLOOK 200 | INFUSION PUMP | FRN | B BRAUN MEDICAL, INC. | OUTLOOK 200 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |