FDA Adverse Event Malfunction Summary report: N

PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)

MDR report key: 4143156 · Received October 6, 2014

Report

Report Number
1031452-2014-12112
Event Type
Malfunction
Date Received
October 6, 2014
Report Date
September 16, 2014
Manufacturer
INVACARE FLORIDA OPERATIONS
Product Code
CAW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Description of Event or Problem · 1

DEALER STATED THAT THE IRC10LXO2 CONCENTRATOR SHUTS DOWN AFTER 30 SECOND, 1 RED LIGHT AND 2 GREEN LIGHTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
623761 PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) 868.5440 CAW INVACARE FLORIDA OPERATIONS IRC10LXO2

Patients

Seq Age Sex Outcome Treatment
1 Other