EON
Report
- Report Number
- 1627487-2014-26771
- Event Type
- Injury
- Date Received
- September 23, 2014
- Date of Event
- September 2, 2014
- Report Date
- September 2, 2014
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- GZB
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS IPG SERIAL NUMBER WAS INCLUDED IN A FIELD CORRECTION. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AN IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE #1 OF 2. REFERENCE MFR REPORT: 1627487-2014-26772. THE PATIENT REPORTS HE EXPERIENCES POCKET HEATING WHILE CHARGING AND IN ADDITION THE IPG BECOMES RED. A NEW MODEL 3726 REPLACEMENT CHARGER WAS SENT TO THE PATIENT WHICH RESOLVED THE ISSUE. ON 08/01/2012 ST JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PATIENTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PATIENTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 591024 | EON | SCS IPG | GZB | ST. JUDE MEDICAL - NEUROMODULATION | 3716 | 119749 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR | Other | IMPLANT DATE:| SCS LEAD: MODEL 3219 |