FDA Adverse Event Malfunction Summary report: N

KANGAROO EPUMP

MDR report key: 4143094 · Received September 25, 2014

Report

Report Number
4143094
Event Type
Malfunction
Date Received
September 25, 2014
Date of Event
November 27, 2013
Report Date
September 25, 2014
Manufacturer
COVIDIEN
Product Code
LZH
Product Problem
Yes
Report Source
User Facility report
Reporter Location
CA, US

Narratives

Description of Event or Problem · 1

NOTED PUMP ROLLER WHEEL MISSING. UPON EXAM, NO VISIBLE DAMAGE TO CASE, WHEEL PLACEMENT IS BEHIND DOOR. ASSUMPTION THAT WHEEL MUST HAVE FALLEN OFF WHEN FEEDING TUBING REMOVED. ASSUMPTION THAT SMALL SCREW THAT HOLDS WHEEL IN PLACE WAS NOT SUFFICIENTLY TIGHTENED. BIOMED REPLACED PUMP WHEEL AND EVALUATED OTHER PUMPS IN STOCK TO ENSURE APPROPRIATE TENSION TO WHEELS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
597340 KANGAROO EPUMP PUMP, INFUSION, ENTERAL LZH COVIDIEN * *

Patients

Seq Age Sex Outcome Treatment
1 *