FDA Adverse Event
Malfunction
Summary report: N
KANGAROO EPUMP
MDR report key: 4143094
·
Received September 25, 2014
Report
- Report Number
- 4143094
- Event Type
- Malfunction
- Date Received
- September 25, 2014
- Date of Event
- November 27, 2013
- Report Date
- September 25, 2014
- Manufacturer
- COVIDIEN
- Product Code
- LZH
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- CA, US
Narratives
Description of Event or Problem · 1
NOTED PUMP ROLLER WHEEL MISSING. UPON EXAM, NO VISIBLE DAMAGE TO CASE, WHEEL PLACEMENT IS BEHIND DOOR. ASSUMPTION THAT WHEEL MUST HAVE FALLEN OFF WHEN FEEDING TUBING REMOVED. ASSUMPTION THAT SMALL SCREW THAT HOLDS WHEEL IN PLACE WAS NOT SUFFICIENTLY TIGHTENED. BIOMED REPLACED PUMP WHEEL AND EVALUATED OTHER PUMPS IN STOCK TO ENSURE APPROPRIATE TENSION TO WHEELS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 597340 | KANGAROO EPUMP | PUMP, INFUSION, ENTERAL | LZH | COVIDIEN | * | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |