FDA Adverse Event Injury Summary report: N

530G INSULIN PUMP

MDR report key: 4143036 · Received September 27, 2014

Report

Report Number
2032227-2014-29957
Event Type
Injury
Date Received
September 27, 2014
Date of Event
August 29, 2014
Report Date
August 29, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OZO
PMA / PMN Number
120010
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAS CONSTANT MOTOR ERROR DURING REWIND DUE TO MOTOR ENCODER SIGNAL OUT OF PHASE. UNABLE TO PERFORM THE DISPLACEMENT TEST DUE TO MOTOR ERROR ALARM. THE INSULIN PUMP HAS SCRATCHED DISPLAY WINDOW, CRACKED BELT CLIP SLOT AND CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER RECEIVED A MOTOR ERROR ALARM. CUSTOMER'S BLOOD GLUCOSE LEVEL AT THE TIME 449MG/DL. ADVISED INSULIN PUMP WOULD BE REPLACED. NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
603569 530G INSULIN PUMP OZO MEDTRONIC MINIMED MMT-751NAS

Patients

Seq Age Sex Outcome Treatment
1 48 YR