FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 4142920
·
Received September 12, 2014
Report
- Report Number
- 1720753-2014-07876
- Event Type
- Malfunction
- Date Received
- September 12, 2014
- Date of Event
- August 12, 2014
- Report Date
- September 12, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE SYSTEM INTERFACE BOARD WAS EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED INTERMITTENT COMMUNICATION FAILED ERROR MESSAGES. THIS ERROR RESULTED IN X-RAYS BEING DISABLED. THIS RESULTED IN A LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 565323 | 9800 | FLUOROSCOPIC X-RAYS | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|