FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 4142920 · Received September 12, 2014

Report

Report Number
1720753-2014-07876
Event Type
Malfunction
Date Received
September 12, 2014
Date of Event
August 12, 2014
Report Date
September 12, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE SYSTEM INTERFACE BOARD WAS EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED INTERMITTENT COMMUNICATION FAILED ERROR MESSAGES. THIS ERROR RESULTED IN X-RAYS BEING DISABLED. THIS RESULTED IN A LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
565323 9800 FLUOROSCOPIC X-RAYS JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1