FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 4142906 · Received September 12, 2014

Report

Report Number
1720753-2014-07877
Event Type
Malfunction
Date Received
September 12, 2014
Date of Event
August 19, 2014
Report Date
September 12, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE POWER SUPPLY WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED INTERMITTENT COMMUNICATION FAILED ERROR MESSAGES. THIS ERROR RESULTS IN X-RAYS BEING DISABLED. THIS RESULTED IN A LOST IMAGING FUNCTIONALITY. THIS COULD RESULT IN A PROCEDURAL DELAY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
565297 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1