FDA Adverse Event
Malfunction
Summary report: N
6800
MDR report key: 4142877
·
Received September 30, 2014
Report
- Report Number
- 1720753-2014-08325
- Event Type
- Malfunction
- Date Received
- September 30, 2014
- Date of Event
- September 5, 2014
- Report Date
- September 30, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE FSE REMOVED, CLEANED AND REINSTALLED ALL OF THE WORKSTATION PCB ASSEMBLIES. THE 5 VDC POWER SUPPLY WAS ALSO EVALUATED, REPLACED AND CALIBRATED TO THE OPTIMAL OPERATING VOLTAGE. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM FAILED TO ALLOW FLUOROSCOPIC EXPOSURES AND EXHIBITED A NON-RECOVERABLE LOSS OF FUNCTIONALITY. NO PATIENT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 607198 | 6800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 6800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |