FDA Adverse Event Malfunction Summary report: N

6800

MDR report key: 4142877 · Received September 30, 2014

Report

Report Number
1720753-2014-08325
Event Type
Malfunction
Date Received
September 30, 2014
Date of Event
September 5, 2014
Report Date
September 30, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE FSE REMOVED, CLEANED AND REINSTALLED ALL OF THE WORKSTATION PCB ASSEMBLIES. THE 5 VDC POWER SUPPLY WAS ALSO EVALUATED, REPLACED AND CALIBRATED TO THE OPTIMAL OPERATING VOLTAGE. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM FAILED TO ALLOW FLUOROSCOPIC EXPOSURES AND EXHIBITED A NON-RECOVERABLE LOSS OF FUNCTIONALITY. NO PATIENT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
607198 6800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 6800

Patients

Seq Age Sex Outcome Treatment
1