FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 4142847
·
Received September 30, 2014
Report
- Report Number
- 1720753-2014-08304
- Event Type
- Malfunction
- Date Received
- September 30, 2014
- Date of Event
- September 4, 2014
- Report Date
- September 30, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE CALIBRATION FILES WERE RELOADED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE MAINFRAME DISPLAYED A COLLIMATOR CAL REQUIRED ERROR MESSAGE ON BOOT UP. THE FIELD ENGINEER NOTED THAT THE COLLIMATOR CONED (CLOSED) DOWN. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.