FDA Adverse Event
Malfunction
Summary report: N
SYMBIQ DUAL CHANNEL
MDR report key: 4142657
·
Received September 26, 2014
Report
- Report Number
- 9615050-2014-05393
- Event Type
- Malfunction
- Date Received
- September 26, 2014
- Date of Event
- September 2, 2014
- Report Date
- September 3, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K110901
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
TESTING AND INVESTIGATION FOUND THAT THE DEVICE HAS A S233 (OVERTEMPERATURE-PSC) MALFUNCTION ALARM CODE IN THE DEVICE HISTORY. THE PROBABLE CAUSE OF THE S233 MALFUNCTION ALARM WAS A BROKEN COOLING FAN. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
THE CUSTOMER CONTACT REPORTED DURING PREVENTIVE MAINTENANCE TESTING AT THE USER FACILITY, THE DEVICE DISPLAYED AN OVER TEMPERATURE ALARM CODE. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTED DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 601047 | SYMBIQ DUAL CHANNEL | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |