FDA Adverse Event Malfunction Summary report: N

8800

MDR report key: 4142629 · Received September 26, 2014

Report

Report Number
1720753-2014-08250
Event Type
Malfunction
Date Received
September 26, 2014
Date of Event
September 10, 2014
Report Date
September 26, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE PROCESSOR ON CONTROL PCB WAS EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT START UP (NO BOOT). THIS RESULTED IN A TOTAL LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
600885 8800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS 8800