FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 4142618 · Received September 26, 2014

Report

Report Number
1720753-2014-08232
Event Type
Malfunction
Date Received
September 26, 2014
Date of Event
September 10, 2014
Report Date
September 26, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE HIGH VOLTAGE CABLE WAS CLEANED AND GREASED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A PROBLEM WITH THE TUBE, AND THAT IT FAILED DURING AN EXAM. THIS RESULTED IN A LOSS OF IMAGING FUNCTIONALITY. THIS COULD RESULT IN THE DELAY OR CANCELLATION OF A PROCEDURE. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
600565 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS 9800