FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 4142597 · Received September 26, 2014

Report

Report Number
1720753-2014-08222
Event Type
Malfunction
Date Received
September 26, 2014
Date of Event
September 11, 2014
Report Date
September 26, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE POWER CABLE CONNECTION IN THE UNINTERRUPTIBLE POWER SUPPLY (UPS) WAS EVALUATED AND REPAIRED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM EXPERIENCED AN IMMEDIATE LOSS OF SYSTEM FUNCTIONALITY AND AN UN-COMMANDED SHUT DOWN. THERE IS NO REPORT OF A PATIENT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
600577 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1