FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 4142591 · Received September 26, 2014

Report

Report Number
1720753-2014-08239
Event Type
Malfunction
Date Received
September 26, 2014
Date of Event
September 11, 2014
Report Date
September 26, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE INTERCONNECT CABLE WAS REPLACED AND THE POWER SUPPLY WAS ADJUSTED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THE SYSTEM EXPERIENCED INTERMITTENT COMMUNICATION FAILED ERROR MESSAGES. THIS RESULTED IN A LOSS OF IMAGING FUNCTIONALITY. THIS COULD RESULT IN A PROCEDURAL DELAY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
600561 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS 9800

Patients

Seq Age Sex Outcome Treatment
1