FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 4142590
·
Received September 26, 2014
Report
- Report Number
- 1720753-2014-08240
- Event Type
- Malfunction
- Date Received
- September 26, 2014
- Date of Event
- September 11, 2014
- Report Date
- September 26, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE INTERCONNECT CABLE AND LEMO CONNECTOR WERE REPLACED. THE BOARDS AND CONNECTORS WERE RESEATED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM INTERMITTENTLY SHUT DOWN. THIS RESULTED IN AN INTERMITTENT, RECOVERABLE LOSS OF IMAGING FUNCTIONALITY. THIS COULD CAUSE PROCEDURAL DELAYS. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.