FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 4142581
·
Received September 26, 2014
Report
- Report Number
- 1720753-2014-08230
- Event Type
- Malfunction
- Date Received
- September 26, 2014
- Date of Event
- September 9, 2014
- Report Date
- September 26, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE POWER CONTROL BOARD WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM WOULD ALARM WHEN PLUGGED IN AND WOULD NOT BOOT UP. THERE IS NO REPORT OF DEATH OR SERIOUS INJURY ASSOCIATED WITH THE COMPLAINT. THIS RESULTED IN A LOSS OF IMAGING FUNCTIONALITY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 601036 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |