LONG NAIL KIT R2.0, TI, LEFT GAMMA3® Ø11X300MM X 125°
Report
- Report Number
- 0009610622-2014-00537
- Event Type
- Injury
- Date Received
- October 6, 2014
- Date of Event
- September 12, 2014
- Report Date
- September 12, 2014
- Manufacturer
- STRYKER TRAUMA KIEL
- Product Code
- HSB
- PMA / PMN Number
- K034002
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- PHYSICIAN
Narratives
REFERRING TO PRODUCT INQUIRY THE LONG NAIL KIT R2.0, TI, LEFT GAMMA3 Ø11X300MM X 125° IS STATED TO BE THE PRIMARY PRODUCT. THE LAG SCREW AND THE LOCKING SCREW ARE CONSIDERED ASSOCIATED PRODUCTS. FAILURES IN MATERIAL OR MANUFACTURING OF THE AFFECTED NAIL KIT RETURNED WERE NOT FOUND. DIMENSIONAL EXAMINATION REVEALED NO DEVIATIONS IN THE RELEVANT UNDAMAGED AREAS OF THE NAIL RETURNED. THE AFFECTED ITEM REPORTED WAS DOCUMENTED AS FAULTLESS PRIOR TO DISTRIBUTION. THUS, WE EXCLUDE DEVIATIONS IN MATERIAL AND MANUFACTURING. THE RECEIVED NAIL IS COMPLETELY BROKEN IN THE WEBS OF THE PROXIMAL DRILL HOLE FOR THE LAG SCREW. ACCORDING TO THE DAMAGE AND THE EVIDENCES OF THE BREAKAGE SURFACES, THE NAIL BROKE IN A FATIGUE FRACTURE DUE TO AXIAL OVERLOAD, WHEREAS IT HAS TO BE ASCERTAINED THAT THE DATE OF IMPLANTATION IS UNKNOWN AND THUS, THE ENTIRE IMPLANTATION DURATION. DRILL MARKS ARE FOUND AT THE LATERAL EDGE OF THE PROXIMAL BORE, IN PARTICULAR AT THE INSIDE OF THE ANTERIOR WEB AT LATERAL. SAME TRACES ARE VISIBLE AT THE DISTAL PORTION OF THE BORE IN THE AREA OF THE MEDIAL EDGE. THE DRILL MARKS ARE PARTLY COVERED BY CONSIDERABLE FRICTION MARKS, WHICH WERE GENERATED BY THE LAG SCREW AFTER THE NAIL BREAKAGE. THE DRILL MARKS WERE GENERATED BY A DEVIATED LAG SCREW STEP DRILL WHICH MOST LIKELY HAD CREATED THE STARTING POINT OF THE FRACTURE (NOTCHING EFFECT) AT THE LATERAL EDGE OF THE ANTERIOR WEB. THE SET SCREW SHOWS A FLATTENING BESIDE THE VERY TIP, WHICH INDICATES THAT THE SCREW HAD NOT BEEN INSERTED UP TO THE INTENDED POSITION RESP. THAT THE VERY TIP DID NOT REACH THE GROUND OF THE LAG SCREW GROOVE DURING THE FIRST STEP OF THE POSITIONING PROCESS. THIS IS SUPPORTED BY MISSING TRACES OF THE SET SCREW ON THE GROUND OF THE SUPERIOR LAG SCREW GROOVE. MATERIAL DEFORMATION (FRICTION MARKS) AT THE MEDIAL / DISTAL AND THE LATERAL / PROXIMAL EDGE OF THE LAG SCREW HOLE INDICATES HIGH TENSION FORCES CAUSED BY HIGH AXIAL LOAD - TRANSMITTED BY THE LAG SCREW, WHICH SUGGESTS MORE THAN PARTIAL WEIGHT BEARING. THE COUNTERPARTS WERE FOUND ON THE LAG SCREW IN THE AREAS WITH FRICTION MARKS WHERE THE EDGES OF THE SLIDING GROOVES HAD BECOME WORN. THE AFFECTED NAIL IS DESIGNED TO WITHSTAND THE NORMAL LOADS DURING THE IMPLANTATION PERIOD, I.E., THE IMPLANT MUST NEITHER BE EXPOSED TO PEAK LOADS NOR TO CONTINUOUS STRESSES. ANOTHER PREREQUISITE FOR A SUCCESSFUL SUPPLY IS UNDISTURBED, NORMAL BONE HEALING. THIS STATE MUST BE ACHIEVED WITHIN A MEDICALLY RECOGNIZED AND BY SCIENTIFIC ANALYSIS CONFIRMED PERIOD (ABOUT 6 MONTHS) IN SUCH WAY, THAT THE BONE STRENGTH ALLOWS SIGNIFICANTLY INCREASING DISCHARGE OF THE TIME-FIXED IMPLANT MATERIAL. IN CASE THAT SUCH A SITUATION DOES NOT OCCUR, EXCEEDING OF THE FATIGUE STRENGTH IS TO BE EXPECTED AND THUS QUITE PREDICTABLE COMPLICATIONS. BASED ON THE ABOVE THE NAIL BREAKAGE IS NOT LINKED TO A DEFICIENCY OF THE DEVICE, BUT IS RATHER CONSIDERED USER RELATED; MOST LIKELY THE CONSIDERABLE INTRAOPERATIVE DAMAGE OF THE GAMMA3 NAIL CAUSED BY A DEVIATED STEP DRILL, HAD LED TO A SIGNIFICANT WEAKENING OF THE NAIL IN THE AREA OF THE LAG SCREW THRU HOLE AND WAS CONTRIBUTED BY PROLONGED OVERLOAD BY THE PATIENT, FOLLOWED BY THE FATIGUE FRACTURE AFTER AN UNKNOWN IMPLANTATION DURATION. NO NON-CONFORMITY WAS IDENTIFIED.
ONCE THE INVESTIGATION HAS BEEN COMPLETED ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.
IT WAS REPORTED THAT PATIENT HAD REVISION OF LEFT HIP. SALES REP REPORTED GAMMA NAIL BROKE AT THE JUNCTION OF THE LAG SCREW. REP REPORTED, "PER PHYSICIAN, PATIENT WALKED TO MAILBOX AND FELT NAIL/HIP "POP". PATIENT WAS REVISED TO A GAMMA 3 NAIL 11X 320 X 120 DEG.
IT WAS REPORTED THAT PATIENT HAD REVISION OF LEFT HIP. SALES REP REPORTED GAMMA NAIL BROKE AT THE JUNCTION OF THE LAG SCREW. REP REPORTED, "PER PHYSICIAN, PATIENT WALKED TO MAILBOX AND FELT NAIL/HIP "POP". PATIENT WAS REVISED TO A GAMMA 3 NAIL 11X 320 X 120 DEG.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 624687 | LONG NAIL KIT R2.0, TI, LEFT GAMMA3® Ø11X300MM X 125° | INTRAMEDULLARY FIXATION ROD | HSB | STRYKER TRAUMA KIEL | K131238 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |